Titration Curve Calculator: pH, Volume & Indicator Choice

Simulate strong, weak and diprotic acid–base titrations drop by drop. Find equivalence points, compare indicators and infer concentration from an endpoint.

#charge balance #diprotic acids #indicators #pH #titration

Use this titration curve calculator to follow the pH of an aqueous sample as a strong acid or base is added. It is useful for chemistry students checking calculations, laboratory staff comparing indicators, and engineers exploring neutralization before a bench test. Choose a strong monoprotic acid, a weak monoprotic acid, a strong monobasic base, a weak monobasic base, or a diprotic acid. The titrant always supplies one equivalent of strong acid or base per mole.

Enter the sample volume and concentration, then the titrant concentration. For weak species, enter the appropriate pKa or pKb at 25 °C. The defaults represent 25 mL of 0.100 mol/L acetic acid titrated with 0.100 mol/L sodium hydroxide. The diprotic defaults represent a closed carbonic-acid system; change both dissociation constants for another acid. All concentrations refer to moles of the original sample compound, rather than equivalents.

Set the burette volume to inspect a point, or reset it to zero and press ▶ to add titrant. The flask changes colour and the bold part of the curve grows as drops are delivered. The delivered volume rounds down to complete drops, with a final partial drop at the playback stop. Reduce the drop volume to inspect a narrow pH jump. The liquid level and colour are schematic, and mixing is assumed instantaneous.

The equivalence point calculator uses reaction stoichiometry. A monoprotic sample has one equivalence volume; a diprotic acid has points at one and two equivalents of added base. Their pH values come from equilibrium, so equivalence does not always mean pH 7. The key-points table also gives half-neutralization values. For a well-behaved weak conjugate pair these approach pKa, or 14 − pKb for a base. Green bands mark pKa ± 1, where adjacent conjugate forms have ratios between 1:10 and 10:1; they do not measure buffer capacity.

Select an indicator to see its transition band against the curve. Compare all four indicators in the table. The percentage is the largest deviation from the selected equivalence volume at either end of the published colour range. The 1% and 5% ratings are screening thresholds, not guarantees of experimental accuracy. In the table, V low and V high are volumes at the lower and upper pH limits; their order reverses for base samples. A dash marks an unavailable boundary, and Outside means the full transition is not crossed. A pH meter is preferable when there is no sharp, suitably placed transition.

To use the titration calculator for an unknown concentration, select the measured-endpoint mode and enter the net titrant volume. For a diprotic acid, identify which equivalence the endpoint represents. This inference assumes the endpoint coincides with that equivalence; indicator bias, blanks and burette uncertainty remain experimental considerations.

The model solves charge balance, dissociation equilibria and water autoionization together at 25 °C. Activity coefficients are ignored: results are ideal dilute-solution estimates. Mixed analytes, initially added salts, weak titrants, gas exchange, precipitation and temperature variation are outside its scope.

How it is calculated

The equilibrium formulation below is derived here from mass conservation, electroneutrality and mass action. OpenStax Chemistry 2e §§14.3, 14.5 and 14.7 provide the equilibrium relations and independent worked examples.

Let h=[H3O+]/c∘h=[H_3O^+]/c^\circ, c∘=1 mol/Lc^\circ=1\,\mathrm{mol/L}, c0=Cs/c∘c_0=C_s/c^\circ, ct=Ct/c∘c_t=C_t/c^\circ, and z=CtV/(CsVs)z=C_tV/(C_sV_s). Dilution gives cA=c0/(1+zc0/ct)c_A=c_0/(1+zc_0/c_t). The normalized spectator-ion concentration is zcAzc_A.

The mean analyte charge q(h)q(h) is −1-1 for a strong monoprotic acid, +1+1 for a strong monobasic base, −Ka/(h+Ka)-K_a/(h+K_a) for HA, and h/(h+Ka,BH+)h/(h+K_{a,BH^+}) for B, with Ka,BH+=Kw/KbK_{a,BH^+}=K_w/K_b. For H₂A:

q(h)=−Ka1(h+2Ka2)h2+Ka1h+Ka1Ka2.q(h)=-\frac{K_{a1}(h+2K_{a2})}{h^2+K_{a1}h+K_{a1}K_{a2}}.

The solver bisects pH on −2…16 for 46 iterations to satisfy

h−Kwh+cA(q(h)+sz)=0,h-\frac{K_w}{h}+c_A\left(q(h)+sz\right)=0,

where s=+1s=+1 for base titrant and −1 for acid titrant. Here Kw=10−14K_w=10^{-14} at 25 °C. Constants and concentrations use the same standard-state normalization. No Henderson–Hasselbalch approximation or excess-reagent shortcut is used for the reported pH.

Stoichiometry gives Veq,j=jCsVs/CtV_{eq,j}=jC_sV_s/C_t and Vhalf,j=(j−1/2)CsVs/CtV_{half,j}=(j-1/2)C_sV_s/C_t. In endpoint mode Cs=CtVendpoint/(jVs)C_s=C_tV_{endpoint}/(jV_s). These are molecular concentrations, with j=1j=1 or 2.

For each indicator boundary, the same charge balance is solved algebraically for zz. Boundaries outside the run are left blank. The comparison reports 100max⁡∣Vboundary/Veq,j−1∣100\max|V_{boundary}/V_{eq,j}-1|. Merck's attached indicator guide supplies the four transition intervals (pp. 6–8). The full curve is cached independently of burette playback.

Assumptions

  • Fixed 25 °C; Kw=10−14K_w=10^{-14}; equilibrium reached after each drop.
  • Activity coefficients are ignored: concentrations replace activities. Volumes add.
  • One initially unsalted analyte: HA, B, H₂A, or a fully dissociated monoprotic acid/monobasic base. Strong titrant contributes one H⁺ or OH⁻ per mole.
  • Spectator ions are inert. No precipitation, complexes, gas exchange or indicator loading.
  • The carbonic-acid defaults use OpenStax's effective constants in a closed system. An open carbonate sample exchanges CO₂ and needs a different model.
  • Flask colours and liquid levels are original schematic illustrations, not optical or vessel-volume predictions. Colour stages follow pH across the supplier's range; exact hues are not calibrated.
  • The 1% and 5% indicator ratings are author-selected screening thresholds for the whole transition interval.

Where it applies

Inputs cover 0.000001–0.5 mol/L; concentrations above 0.2 mol/L trigger a dilute-model warning. Activity effects can matter even below that threshold, particularly for multivalent ions. For accurate analytical work use suitable activities, standardized titrant, a validated endpoint and measured uncertainty.

Diprotic pKa₂ must exceed pKa₁. The coupled equations also handle overlapping dissociations, but separate stoichiometric points may not produce distinct jumps. Half-neutralization pH need not equal pKa when water, dilution or overlapping equilibria matter. pKa ± 1 bands show species ratios, not absolute buffering strength.

Burette playback stops at twice the final stoichiometric equivalence volume. The selected volume is rounded down to complete drops, except for the final partial drop. Indicator comparisons use continuous equilibrium volumes, so their resolution can exceed the chosen drop size. A blank transition boundary means it is outside this run.

Endpoint concentration assumes the measured endpoint is the selected equivalence. This is a stoichiometric inference, not a correction for indicator bias or a titration-curve fit. No weak titrants, mixtures, triprotic samples, initially added salts, variable temperature or gas-transfer kinetics are modeled.

References

  • OpenStax, Chemistry 2e, §14.7 Acid–Base Titrations, Examples 14.21–14.22 and Table 14.2. https://openstax.org/books/chemistry-2e/pages/14-7-acid-base-titrations
  • OpenStax, Chemistry 2e, §14.5 Polyprotic Acids, Example 14.19 (carbonic acid constants and concentrations). https://openstax.org/books/chemistry-2e/pages/14-5-polyprotic-acids
  • OpenStax, Chemistry 2e, §4.5 Quantitative Chemical Analysis, Example 4.14 (endpoint concentration). https://openstax.org/books/chemistry-2e/pages/4-5-quantitative-chemical-analysis
  • OpenStax, Chemistry 2e, §14.3 Relative Strengths of Acids and Bases (weak-base equilibrium and ammonia example). https://openstax.org/books/chemistry-2e/pages/14-3-relative-strengths-of-acids-and-bases
  • Merck / MilliporeSigma, Colorize Your Lab: Reliable indicators for analytical titrations, pp. 6–8. https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/marketing/global/documents/106/158/11689en-colorize-your-lab-ms.pdf

Inputs

InputSymbolUnitDefault
Calculation
Known concentration predicts a titration curve. Measured endpoint infers the sample concentration from titrant delivery at the selected equivalence.
calc_modeKnown concentration
Sample chemistry
Acid models are monoprotic except for H₂A; bases accept one proton per molecule. The diprotic sample starts as H₂A, without added salts. Use a strong titrant supplying one H⁺ or OH⁻ per mole.
systemWeak acid + strong base
Sample volumeV_smL25
Sample concentrationC_knownmol/L0.1
Acid pKa
Default: acetic acid, Ka = 1.8 × 10⁻⁵ at 25 °C. Enter −log₁₀(Ka).
pKa4.744727494896694
Base pKb
Default: ammonia, Kb = 1.8 × 10⁻⁵ at 25 °C. Enter −log₁₀(Kb).
pKb4.744727494896694
Diprotic pKa₁
Default pair: closed carbonic-acid system, Ka₁ = 4.3 × 10⁻⁷ and Ka₂ = 4.7 × 10⁻¹¹.
pKa_d16.366531544420414
Diprotic pKa₂pKa_d210.327902142064282
Strong titrant concentrationC_tmol/L0.1
Equivalence to assess
First: 1 mol OH⁻ per mol H₂A. Second: 2 mol OH⁻ per mol H₂A. Used for indicator comparison and endpoint concentration. Two stoichiometric points do not guarantee two distinct pH jumps.
eq_stageSecond equivalence
Measured endpoint volume
Enter the net burette delivery after any blank correction. Concentration assumes this endpoint represents the selected equivalence.
V_endpointmL25
IndicatorindicatorPhenolphthalein
Volume per drop
Illustrative drop size. The delivered volume and pH advance by complete drops; choose a smaller drop for finer resolution.
drop_volumemL0.05
Burette volume — ▶ to add drops
Set to zero before playing a full run. Playback stops at twice the final equivalence volume. The result uses complete drops, including a final partial drop at the stop.
V_addedmL25

Results

With the default inputs above, the calc gives:

ResultValue
Flask pH8.722
Delivered titrant25 mL
Selected equivalence volume25 mL
pH at selected equivalence8.722
Sample concentration0.1 mol/L
pH at selected half-neutralization4.745
Indicator assessmentWithin 1%: the full colour range lies close to selected equivalence.
Current conjugate-pair regionOutside the pKa ± 1 conjugate-pair bands.
Selected indicator transitionFull transition: 24.992–25.031 mL
Worst indicator volume deviation (%)0.1254

Checked against sources

Each case below was calculated by this calc and agrees with the value from its source (within the tolerance where one is shown).

HCl / NaOH equivalence

Source: OpenStax Chemistry 2e §14.7 Example 14.21(c): 25.00 mL 0.100 M HCl + 25.00 mL 0.100 M NaOH; pH 7.00. https://openstax.org/books/chemistry-2e/pages/14-7-acid-base-titrations

Inputs: Sample chemistry = Strong acid + strong base; the others as above.

ResultSourceThis calc
Flask pH77
Selected equivalence volume25 mL25 mL

Acetic acid / NaOH equivalence

Source: OpenStax Chemistry 2e §14.7 Example 14.22(b): 25.00 mL 0.100 M acetic acid, Ka = 1.8e-5, and 25.00 mL 0.100 M NaOH; pH 8.72 (rounded, hydrolysis approximation). https://openstax.org/books/chemistry-2e/pages/14-7-acid-base-titrations

Inputs: the default inputs above.

ResultSourceThis calc
Flask pH8.72 ± 0.018.722

Strong acid: start, half, equivalence and excess titrant

Source: OpenStax Chemistry 2e §14.7 Example 14.21: 25.00 mL 0.100 M HCl, 0.100 M NaOH; 0.00, 12.50, 25.00 and 37.50 mL. Half-stage pH is −log₁₀(0.0333) using the example's stated excess concentration (rounded). https://openstax.org/books/chemistry-2e/pages/14-7-acid-base-titrations

Inputs: Sample chemistry = Strong acid + strong base; Burette volume — ▶ to add drops = 37.5 mL; the others as above.

ResultSourceThis calc
Initial pH11
pH at selected half-neutralization1.4771.477
pH at selected equivalence77
Flask pH12.312.301

Acetic acid: initial, half and post-equivalence

Source: OpenStax Chemistry 2e §14.7 Example 14.22: 25.00 mL 0.100 M acetic acid, Ka 1.8e-5, with 0.100 M NaOH; reported pH 2.87 initially, 4.74 at 12.50 mL, and 12.30 at 37.50 mL. The 0.01 tolerance admits textbook rounding and small-ionization approximations. https://openstax.org/books/chemistry-2e/pages/14-7-acid-base-titrations

Inputs: Burette volume — ▶ to add drops = 37.5 mL; the others as above.

ResultSourceThis calc
Initial pH2.87 ± 0.012.875
pH at selected half-neutralization4.74 ± 0.014.745
Flask pH12.312.301

Diprotic carbonic acid: independent equilibrium example

Source: OpenStax Chemistry 2e §14.5 Example 14.19: initial H₂CO₃ = 0.033 M; Ka₁ = 4.3e-7, Ka₂ = 4.7e-11; [H₃O⁺] = 1.2e-4 M (two significant digits, first-step small-ionization approximation). The simultaneous charge balance retains both ionizations; tolerance covers rounding. https://openstax.org/books/chemistry-2e/pages/14-5-polyprotic-acids

Inputs: Sample chemistry = Diprotic acid + strong base; Sample concentration = 0.033 mol/L; Burette volume — ▶ to add drops = 0 mL; the others as above.

ResultSourceThis calc
Hydronium concentration1.2 × 10⁻⁴ ± 2 × 10⁻⁶ mol/L1.189 × 10⁻⁴ mol/L

Weak base: published ammonia hydroxide concentration

Source: OpenStax Chemistry 2e §14.3, Check Your Learning after Example 14.13: 0.0325 M NH₃, Kb=1.76e-5, answer [OH⁻]=7.56e-4 M from the small-ionization approximation. The exact depletion solution is slightly lower; absolute tolerance 1e-5 M covers that approximation without changing the physics. https://openstax.org/books/chemistry-2e/pages/14-3-relative-strengths-of-acids-and-bases

Inputs: Sample chemistry = Weak base + strong acid; Sample concentration = 0.0325 mol/L; Base pKb = 4.754; Burette volume — ▶ to add drops = 0 mL; Indicator = Methyl red; the others as above.

ResultSourceThis calc
Hydroxide concentration7.56 × 10⁻⁴ ± 1 × 10⁻⁵ mol/L7.476 × 10⁻⁴ mol/L

Unknown HCl concentration from a measured endpoint

Source: OpenStax Chemistry 2e §4.5 Example 4.14: 50.00 mL HCl, endpoint 35.23 mL of 0.250 M NaOH, reported HCl molarity 0.176 M. This tests molecular stoichiometry; it is not a validation of ideal-solution pH at 0.250 M. https://openstax.org/books/chemistry-2e/pages/4-5-quantitative-chemical-analysis

Inputs: Calculation = Measured endpoint; Sample chemistry = Strong acid + strong base; Sample volume = 50 mL; Strong titrant concentration = 0.25 mol/L; Measured endpoint volume = 35.23 mL; the others as above.

ResultSourceThis calc
Sample concentration0.176 ± 5 × 10⁻⁴ mol/L0.1761 mol/L
Selected equivalence volume35.23 mL35.23 mL

Supplier phenolphthalein transition range

Source: Merck Colorize Your Lab, p. 8, phenolphthalein indicator ACS / Reag. Ph Eur: pH 8.2–9.8. Checks transfer of supplier data, not the pH method. https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/marketing/global/documents/106/158/11689en-colorize-your-lab-ms.pdf

Inputs: the default inputs above.

ResultSourceThis calc
Selected indicator lower pH limit8.28.2
Selected indicator upper pH limit9.89.8